Huazheng HZKS-3 ASTM D92 penskymartens flash point apparatus Open Cup Flash Point Tester

Safety Instructions
1.Be sure to use a grounded outlet for the instrument’s power supply.
2.No touching the Oil Cups and its nearby parts during test, in case of scald.
3.There must be someone on duty during the test. Cover oil cup lid in time if the sample is burning after the test is completed.
4.No touch the oil lid because of high heat.
5.Do not touch the ignition rod by hands; otherwise it will cause permanent damage to the instrument!
Instrument Features
HZKS-3, with touch screen to replace the keyboard operation, is used to measure the open flash point value of petroleum products. Using advanced technology abroad, LCD screen full English display man-machine dialogue interface, full screen touch keys, convenient and fast, open beam, fuzzy control integrated software, modular structure.
1.Standard modular design, one main engine can control the number of open or closed test furnaces, be freely exchanged between the test furnace, According to user needs, greatly improving the efficiency and economic benefits.
2.High-speed signal processor controlled, high-precision analog / digital converter circuit, the instrument is reliable and high precision.
3.Auto-ignition.
4.Automatic measurement of atmospheric pressure and automatic correction.
5.Adaptive PID control algorithm to automatically adjust the temperature according to standard curve.
6.The whole experiment process is completed automatically, including detection, opening the cover, ignition, alarm, cooling and printing.
7.Automatically stop detection and alarm when the temperature is out of range.
8.Thermal micro-printer, making printing more efficient and more beautiful, with off-line printing.
9.History records with time stamp, up to 256.
10.Large screen with 320x240 graphical LCD display, showing a rich, full character display.
11.touch screen operation, friendly interface; The screen saver function, extend the life of the LCD.
12.Interface, RS232 interface for printer or downloads on PC.
Technical Parameters
1.Temperature Measurement
Temperature measuring range: -59.9°C~399.9°C
Temperature Control Range: Room temperature~350°C
Repeatability: ≤150℃ 3℃
>150℃ 5℃
Resolution: 0.1°C
Precision: 0.5%
Temperature Sensor: Platinum Resistor(PT100)
Flash Fire Sensor: Ion detection ring
2.Ambient Temperature: 0 ~ 40°C
3.Relative Humidity: ≤ 85% RH
4.Power Supply: AC220V ± 20% ; 50HZ ± 10%
5.Power Consumption: approx 400W
6.Overall Dimension: 440X280X290(mm)
7.Instrument Weight: 14kg





Precautions And Instrument Maintenance
(1)The instrument should be used in a non-corrosive environment.
(2)The oil cup should be cleaned, when replacing test sample.
(3)Check thermocouple should not have oil contamination, or it will affect the detection sensitivity.
(4)No touch ignition head, oil cup and its surrounding parts to avoid scorching during the test.
Failure And Treatment
NO. |
Failure |
Treatment |
1 | Turn on the power, the LCD screen no display | Check power supply is plugged in, open the instrument measured the board examine the plug is not loose. |
2 |
Large repeatability error for sample testing | Remove the lifting arm top cover, and check 1. whether the mixer shaft is broken 2. Oil may effect on thermocouple sensitivity, plz dry it by filter paper |
3 |
No heating | Electric wire broken |
4 | Lift arm rise or fall failure | Limit opt coupler damage The screw between main engine and lift arm fell off |
Packing List
No. |
Item |
Qty |
1 |
Main engine |
1 |
2 |
Test cup |
1 |
3 |
Fuse pipe |
2 |
4 |
Power line |
1 |
5 |
Touch pen |
1 |
6 |
Fire cover |
1 |
7 |
Print paper |
1 |